Our NG-P series PSA nitrogen generator produce high-purity N₂ (95%–99.999%) from compressed air using carbon molecular sieve (CMS) and alternating pressure cycles. Ideal for labs needing 99.99%+ purity (e.g.,GC, LC-MS), they outperform membrane units in purity and stability.

Core Principle of PSA nitrogen generator

1.Compressed Air Pretreatment:Air is compressed (7–8 bar), dried, and filtered to remove water, oil, and particles.

2.Dual-Tower PSA Cycle (A/B Towers)
Adsorption (High Pressure, ~7 bar)
Compressed air enters a tower filled with CMS. Smaller O₂, CO₂, and moisture are adsorbed; larger N₂ passes through as product gas.
Regeneration (Low Pressure, ~1 bar)
The other tower depressurizes, releasing adsorbed O₂ to atmosphere. Towers alternate every 60–120 seconds for continuous flow.

PRODUCT SPECIFICATIONS

ModelPNG500
Nitrogen flow300ml/min
Nitrogen purity99.999%
Pressure4bar
Outlet size1/8″ BSPP
VoltageAC220v,50Hz