Introduction to ZX series zero air generator
The ZX series zero gas generator is a device specifically designed for producing “Zero Air”, which refers to high-purity air that is almost free of hydrocarbons, carbon dioxide, water vapor, particulate matter, and other impurities that may interfere with analysis (typically<0.15 ppm HC). It is widely used for baseline calibration and combustion gas supply in laboratories, process analysis, and on-site gas detectors.
Application
- Gas chromatography (GC): provides combustion gas or baseline zero gas for detectors such as FID, NPD, FPD, PFPD, significantly reducing background noise and improving detection sensitivity.
- Environmental monitoring instrument calibration: used for zero point calibration of on-site instruments such as infrared gas analyzers (IRGA), eliminating the need for high-pressure gas cylinders.
- Process analysis: Provide zero point reference gas in industrial process control to ensure the accuracy of continuous online monitoring.
- General laboratory use: Provide a stable gas source for experiments that require dry, low impurity air, such as combustion experiments and gas mixing.

Parameter
| Model | ZX5000 |
| Max output flow | 5L/min |
| Output THC content | <0.1ppm |
| Output pressure | 0.1-0.6Mpa (without built-in air compressor, depending on inlet air pressure) |
| Dew point | <-40℃(also have relation to input air dew point) |
| Particle | <0.01um |
| Input air THC content | ﹤100ppm |
| Power supply | AC220v±10%,50-60Hz |
| Power | 300w |
| Instrument start time before stable | ﹤1h |
| Enviroment temperature | 10-40℃ |
| Heating furnace operating temperature | 350℃ |
| Size | 400×430×177(mm) |